SPECIFICATIONS AND PERFORMANCE
ELECTROMAGNETIC COMPATIBILITY
The product conforms to the EMC standard (IEC 60601-1-2 Ed2.1:2004)
a)MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in the
ACCOMPANYING DOCUMENTS.
b)Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c)The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYS-
TEM as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d)The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equip-
ment. If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be
observed to verify normal operation in the configuration in which it will be used.
e)The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other
than those specified may result in increased EMISSION or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
Item Parts cord Length(m)
AC power cord (for test device) 44804 70020 1.5
RS232C cable 41201 53100 1.4
RS232C cable 41201 53100 1.4
USB cable 3.0
Guidance and manufacturer's declaration - electromagnetic emissions
The RM-8900,KR-8900 is intended for use in the electromagnetic environment specified below.
The customer or the user of the RM-8900,KR-8900 should assure that it is used in such an
environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The RM-8900,KR-8900 uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any inter-
ference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The RM-8900,KR-8900 is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies